| 48 T | $265.00 | 
| 96 T | $376.00 | 
| 96 T*5 | $1598.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 20000.00 | 0.191 | |
| 10000.00 | 0.352 | |
| 5000.00 | 0.577 | |
| 2500.00 | 0.833 | |
| 1250.00 | 1.197 | |
| 625.00 | 1.578 | |
| 312.50 | 1.789 | |
| 0.00 | 2.123 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 80-96% | 88% | 
| EDTA plasma(n=5) | 80-92% | 86% | 
| Heparin plasma(n=5) | 78-95% | 87% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 88-97% | 95-106% | 87-98% | 93-102% | 
| EDTA plasma(n=5) | 86-97% | 93-102% | 87-98% | 83-97% | 
| Heparin plasma(n=5) | 95-103% | 87-98% | 85-92% | 79-96% |