| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 500.00 | 1.961 | 1.875 |
| 250.00 | 1.647 | 1.561 |
| 125.00 | 1.194 | 1.108 |
| 62.50 | 0.831 | 0.745 |
| 31.25 | 0.496 | 0.410 |
| 15.63 | 0.381 | 0.295 |
| 7.82 | 0.129 | 0.043 |
| 0.00 | 0.086 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 95-107% | 101% |
| EDTA plasma(n=5) | 96-104% | 100% |
| Heparin plasma(n=5) | 98-104% | 102% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-94% | 93-101% | 87-98% | 89-102% |
| EDTA plasma(n=5) | 87-94% | 85-96% | 89-100% | 85-96% |
| Heparin plasma(n=5) | 82-93% | 95-104% | 82-96% | 85-98% |