| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10000.00 | 0.185 | |
| 5000.00 | 0.376 | |
| 2500.00 | 0.577 | |
| 1250.00 | 0.833 | |
| 625.00 | 1.119 | |
| 312.50 | 1.522 | |
| 156.25 | 1.762 | |
| 0.00 | 2.115 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 92-106% | 99% |
| EDTA plasma(n=5) | 88-103% | 95% |
| Heparin plasma(n=5) | 77-95% | 91% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 79-96% | 85-99% | 87-105% | 87-99% |
| EDTA plasma(n=5) | 85-96% | 87-101% | 89-102% | 85-98% |
| Heparin plasma(n=5) | 92-106% | 92-101% | 80-94% | 86-101% |