| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 800.00 | 2.195 | 2.091 |
| 400.00 | 1.579 | 1.475 |
| 200.00 | 1.197 | 1.093 |
| 100.00 | 0.826 | 0.722 |
| 50.00 | 0.563 | 0.459 |
| 25.00 | 0.369 | 0.265 |
| 12.50 | 0.226 | 0.122 |
| 0.00 | 0.104 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-99% | 93% |
| EDTA plasma(n=5) | 92-105% | 98% |
| Heparin plasma(n=5) | 83-95% | 101% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 86-95% | 93-103% | 86-97% | 90-98% |
| EDTA plasma(n=5) | 91-99% | 83-89% | 88-94% | 87-93% |
| Heparin plasma(n=5) | 89-97% | 95-104% | 82-96% | 85-98% |