| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 500.00 | 1.981 | 1.898 | 
| 250.00 | 1.565 | 1.482 | 
| 125.00 | 1.193 | 1.110 | 
| 62.50 | 0.851 | 0.768 | 
| 31.25 | 0.475 | 0.392 | 
| 15.63 | 0.249 | 0.166 | 
| 7.81 | 0.152 | 0.069 | 
| 0.00 | 0.083 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 86-115% | 101% | 
| EDTA plasma(n=5) | 80-109% | 95% | 
| Heparin plasma(n=5) | 81-110% | 96% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 85-111% | 85-103% | 88-99% | 84-101% | 
| EDTA plasma(n=5) | 89-119% | 88-106% | 90-115% | 82-90% | 
| Heparin plasma(n=5) | 92-116% | 94-117% | 85-100% | 91-100% | 