| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.136 | 2.053 |
| 5.00 | 1.532 | 1.449 |
| 2.50 | 1.147 | 1.064 |
| 1.25 | 0.878 | 0.795 |
| 0.63 | 0.543 | 0.460 |
| 0.32 | 0.324 | 0.241 |
| 0.16 | 0.226 | 0.143 |
| 0.00 | 0.083 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 81-95% | 88% |
| EDTA plasma(n=5) | 94-105% | 98% |
| Heparin plasma(n=5) | 81-95% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-102% | 92-101% | 88-102% | 87-102% |
| EDTA plasma(n=5) | 85-101% | 93-104% | 87-103% | 91-102% |
| Heparin plasma(n=5) | 89-97% | 85-101% | 84-98% | 85-93% |