| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 10000.00 | 0.175 | |
| 5000.00 | 0.367 | |
| 2500.00 | 0.621 | |
| 1250.00 | 0.861 | |
| 625.00 | 1.197 | |
| 312.50 | 1.578 | |
| 156.25 | 1.799 | |
| 0.00 | 2.132 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 94-106% | 100% |
| EDTA plasma(n=5) | 80-97% | 88% |
| Heparin plasma(n=5) | 77-95% | 91% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 86-94% | 85-95% | 87-99% | 99-105% |
| EDTA plasma(n=5) | 89-98% | 83-103% | 98-105% | 92-101% |
| Heparin plasma(n=5) | 87-96% | 86-98% | 92-103% | 87-93% |