48 T | $320.00 |
96 T | $458.00 |
Concentration (pg/mL) | OD | Corrected OD |
---|---|---|
10000.00 | 0.175 | |
5000.00 | 0.367 | |
2500.00 | 0.621 | |
1250.00 | 0.861 | |
625.00 | 1.197 | |
312.50 | 1.578 | |
156.25 | 1.799 | |
0.00 | 2.132 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 94-106% | 100% |
EDTA plasma(n=5) | 80-97% | 88% |
Heparin plasma(n=5) | 77-95% | 91% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 86-94% | 85-95% | 87-99% | 99-105% |
EDTA plasma(n=5) | 89-98% | 83-103% | 98-105% | 92-101% |
Heparin plasma(n=5) | 87-96% | 86-98% | 92-103% | 87-93% |