Products

Human INPP4B(Inositol Polyphosphate-4-Phosphatase Type II 105kDa) ELISA Kit
One-click to copy product information
ELK7595
Size: Price:
48 T $320.00
96 T $458.00

Overview

Product name: Human INPP4B(Inositol Polyphosphate-4-Phosphatase Type II 105kDa) ELISA Kit
Reactivity: Human
Alternative Names: Type II inositol 3,4-bisphosphate 4-phosphatase
Assay Type: Sandwich
Sensitivity: 0.67 ng/mL
Standard: 100 ng/mL
Detection Range: 1.57-100 ng/mL
Sample type: Tissue homogenates, cell lysates and other biological fluids
Assay length: 3.5h
Research Area: Enzyme & Kinase;
Uniprot ID: O15327
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human INPP4B. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human INPP4B. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human INPP4B, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human INPP4B in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (ng/mL) OD Corrected OD
100.00 2.035 1.942
50.00 1.571 1.478
25.00 1.167 1.074
12.50 0.878 0.785
6.25 0.457 0.364
3.13 0.366 0.273
1.57 0.134 0.041
0.00 0.093 0.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant INPP4B and the recovery rates were calculated by comparing the measured value to the expected amount of INPP4B in samples.
Matrix Recovery range Average
serum(n=5) 78-96% 87%
EDTA plasma(n=5) 80-95% 87%
Heparin plasma(n=5) 78-92% 85%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of INPP4B and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 78-93% 96-102% 95-103% 89-97%
EDTA plasma(n=5) 85-92% 83-96% 93-102% 87-98%
Heparin plasma(n=5) 92-103% 93-105% 88-101% 97-106%