| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.059 | 1.962 |
| 5.00 | 1.328 | 1.231 |
| 2.50 | 1.034 | 0.937 |
| 1.25 | 0.828 | 0.731 |
| 0.63 | 0.559 | 0.462 |
| 0.32 | 0.395 | 0.298 |
| 0.16 | 0.241 | 0.144 |
| 0.00 | 0.097 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-99% | 93% |
| EDTA plasma(n=5) | 93-107% | 100% |
| Heparin plasma(n=5) | 85-99% | 92% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-96% | 86-95% | 90-102% | 81-90% |
| EDTA plasma(n=5) | 79-91% | 95-104% | 93-102% | 79-94% |
| Heparin plasma(n=5) | 87-98% | 92-104% | 82-96% | 95-102% |