| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 40.00 | 2.169 | 2.075 |
| 20.00 | 1.647 | 1.553 |
| 10.00 | 1.151 | 1.057 |
| 5.00 | 0.891 | 0.797 |
| 2.50 | 0.496 | 0.402 |
| 1.25 | 0.415 | 0.321 |
| 0.63 | 0.211 | 0.117 |
| 0.00 | 0.094 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-97% | 88% |
| EDTA plasma(n=5) | 78-92% | 85% |
| Heparin plasma(n=5) | 90-105% | 97% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 98-103% | 86-92% | 87-93% | 93-102% |
| EDTA plasma(n=5) | 85-92% | 79-96% | 89-101% | 85-97% |
| Heparin plasma(n=5) | 86-94% | 93-102% | 89-97% | 85-94% |