| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 10.00 | 2.235 | 2.147 | 
| 5.00 | 1.557 | 1.469 | 
| 2.50 | 1.108 | 1.020 | 
| 1.25 | 0.861 | 0.773 | 
| 0.63 | 0.513 | 0.425 | 
| 0.32 | 0.305 | 0.217 | 
| 0.16 | 0.187 | 0.099 | 
| 0.00 | 0.088 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 87-99% | 93% | 
| EDTA plasma(n=5) | 92-105% | 98% | 
| Heparin plasma(n=5) | 83-95% | 89% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 87-98% | 92-104% | 82-96% | 95-102% | 
| EDTA plasma(n=5) | 82-93% | 95-104% | 82-96% | 85-98% | 
| Heparin plasma(n=5) | 95-104% | 82-96% | 92-105% | 83-92% |