| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 2.270 | 2.175 |
| 2500.00 | 1.621 | 1.526 |
| 1250.00 | 1.167 | 1.072 |
| 625.00 | 0.878 | 0.783 |
| 312.50 | 0.536 | 0.441 |
| 156.25 | 0.262 | 0.167 |
| 78.13 | 0.225 | 0.130 |
| 0.00 | 0.095 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 86-98% | 92% |
| EDTA plasma(n=5) | 95-107% | 101% |
| Heparin plasma(n=5) | 83-95% | 89% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-98% | 83-103% | 98-105% | 92-101% |
| EDTA plasma(n=5) | 88-97% | 79-92% | 85-93% | 89-97% |
| Heparin plasma(n=5) | 85-94% | 97-103% | 82-90% | 92-101% |