| 48 T | $412.00 | 
| 96 T | $588.00 | 
| 96 T*5 | $2499.00 | 
| Concentration (mIU/mL) | OD | Corrected OD | 
|---|---|---|
| 100.00 | 2.154 | 2.059 | 
| 50.00 | 1.686 | 1.591 | 
| 25.00 | 1.162 | 1.067 | 
| 12.50 | 0.834 | 0.739 | 
| 6.25 | 0.527 | 0.432 | 
| 3.13 | 0.369 | 0.274 | 
| 1.57 | 0.235 | 0.140 | 
| 0.00 | 0.095 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 84-98% | 91% | 
| EDTA plasma(n=5) | 79-97% | 88% | 
| Heparin plasma(n=5) | 80-95% | 97% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 85-96% | 87-101% | 93-99% | 87-101% | 
| EDTA plasma(n=5) | 88-100% | 95-103% | 89-97% | 85-94% | 
| Heparin plasma(n=5) | 88-97% | 79-92% | 97-105% | 89-97% |