| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 200.00 | 2.168 | 2.076 | 
| 100.00 | 1.700 | 1.608 | 
| 50.00 | 1.269 | 1.177 | 
| 25.00 | 0.889 | 0.797 | 
| 12.50 | 0.649 | 0.557 | 
| 6.25 | 0.342 | 0.250 | 
| 3.13 | 0.283 | 0.191 | 
| 0.00 | 0.092 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 92-105% | 98% | 
| EDTA plasma(n=5) | 84-96% | 90% | 
| Heparin plasma(n=5) | 80-95% | 87% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 93-105% | 88-102% | 87-98% | 85-96% | 
| EDTA plasma(n=5) | 93-105% | 82-101% | 91-103% | 95-102% | 
| Heparin plasma(n=5) | 88-104% | 84-98% | 95-105% | 88-102% |