| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 3000.00 | 0.185 | |
| 1500.00 | 0.352 | |
| 750.00 | 0.557 | |
| 375.00 | 0.857 | |
| 187.50 | 1.152 | |
| 93.75 | 1.565 | |
| 46.88 | 1.799 | |
| 0.00 | 2.132 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 85-97% | 91% | 
| EDTA plasma(n=5) | 96-107% | 101% | 
| Heparin plasma(n=5) | 90-105% | 93% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 85-95% | 85-96% | 87-98% | 87-100% | 
| EDTA plasma(n=5) | 88-99% | 81-93% | 89-97% | 85-94% | 
| Heparin plasma(n=5) | 95-103% | 86-92% | 79-103% | 89-97% |