| 48 T | $412.00 | 
| 96 T | $588.00 | 
| 96 T*5 | $2499.00 | 
| Concentration (mIU/mL) | OD | Corrected OD | 
|---|---|---|
| 100.00 | 1.948 | 1.853 | 
| 50.00 | 1.663 | 1.568 | 
| 25.00 | 1.202 | 1.107 | 
| 12.50 | 0.876 | 0.781 | 
| 6.25 | 0.545 | 0.450 | 
| 3.13 | 0.322 | 0.227 | 
| 1.57 | 0.215 | 0.120 | 
| 0.00 | 0.095 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 82-98% | 90% | 
| EDTA plasma(n=5) | 90-107% | 98% | 
| Heparin plasma(n=5) | 87-99% | 96% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 85-92% | 79-96% | 92-101% | 87-98% | 
| EDTA plasma(n=5) | 95-103% | 86-92% | 89-97% | 91-103% | 
| Heparin plasma(n=5) | 90-97% | 79-93% | 91-98% | 89-102% |