| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 2000.00 | 1.861 | 1.773 | 
| 1000.00 | 1.497 | 1.409 | 
| 500.00 | 1.216 | 1.128 | 
| 250.00 | 0.752 | 0.664 | 
| 125.00 | 0.586 | 0.498 | 
| 62.50 | 0.354 | 0.266 | 
| 31.25 | 0.196 | 0.108 | 
| 0.00 | 0.088 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 82-95% | 88% | 
| EDTA plasma(n=5) | 90-103% | 96% | 
| Heparin plasma(n=5) | 85-97% | 91% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 89-97% | 85-101% | 84-98% | 85-93% | 
| EDTA plasma(n=5) | 83-95% | 88-102% | 85-97% | 80-92% | 
| Heparin plasma(n=5) | 98-105% | 92-101% | 85-105% | 87-98% |