| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 5000.00 | 1.954 | 1.871 | 
| 2500.00 | 1.638 | 1.555 | 
| 1250.00 | 1.243 | 1.160 | 
| 625.00 | 0.847 | 0.764 | 
| 312.50 | 0.567 | 0.484 | 
| 156.25 | 0.357 | 0.274 | 
| 78.13 | 0.196 | 0.113 | 
| 0.00 | 0.083 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 86-99% | 92% | 
| EDTA plasma(n=5) | 83-98% | 90% | 
| Heparin plasma(n=5) | 80-95% | 87% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 87-98% | 92-105% | 95-107% | 82-96% | 
| EDTA plasma(n=5) | 93-102% | 81-96% | 93-106% | 82-95% | 
| Heparin plasma(n=5) | 86-95% | 93-103% | 87-98% | 85-96% |