| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 2.126 | 2.033 |
| 1000.00 | 1.583 | 1.490 |
| 500.00 | 1.253 | 1.160 |
| 250.00 | 0.780 | 0.687 |
| 125.00 | 0.614 | 0.521 |
| 62.50 | 0.357 | 0.264 |
| 31.25 | 0.240 | 0.147 |
| 0.00 | 0.093 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 90-105% | 97% |
| EDTA plasma(n=5) | 80-95% | 87% |
| Heparin plasma(n=5) | 86-99% | 92% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 95-106% | 87-101% | 82-97% | 87-102% |
| EDTA plasma(n=5) | 92-105% | 82-92% | 86-94% | 81-95% |
| Heparin plasma(n=5) | 95-104% | 82-94% | 86-96% | 81-97% |