| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 5000.00 | 2.154 | 2.085 | 
| 2500.00 | 1.593 | 1.524 | 
| 1250.00 | 0.935 | 0.866 | 
| 625.00 | 0.752 | 0.683 | 
| 312.50 | 0.532 | 0.463 | 
| 156.25 | 0.324 | 0.255 | 
| 78.13 | 0.234 | 0.165 | 
| 0.00 | 0.069 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 80-93% | 86% | 
| EDTA plasma(n=5) | 80-95% | 87% | 
| Heparin plasma(n=5) | 95-107% | 101% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 86-97% | 90-98% | 80-93% | 92-105% | 
| EDTA plasma(n=5) | 86-99% | 93-107% | 85-94% | 87-101% | 
| Heparin plasma(n=5) | 86-101% | 86-101% | 87-96% | 92-105% |