| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 4000.00 | 2.045 | 1.959 | 
| 2000.00 | 1.537 | 1.451 | 
| 1000.00 | 1.087 | 1.001 | 
| 500.00 | 0.782 | 0.696 | 
| 250.00 | 0.509 | 0.423 | 
| 125.00 | 0.336 | 0.250 | 
| 62.50 | 0.214 | 0.128 | 
| 0.00 | 0.086 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 92-106% | 99% | 
| EDTA plasma(n=5) | 77-95% | 86% | 
| Heparin plasma(n=5) | 88-103% | 95% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 88-97% | 97-105% | 81-93% | 89-102% | 
| EDTA plasma(n=5) | 86-97% | 93-105% | 82-93% | 95-104% | 
| Heparin plasma(n=5) | 86-99% | 93-107% | 85-94% | 87-101% |