| 48 T | $412.00 | 
| 96 T | $588.00 | 
| 96 T*5 | $2499.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 1000.00 | 2.207 | 2.111 | 
| 500.00 | 1.629 | 1.533 | 
| 250.00 | 1.073 | 0.977 | 
| 125.00 | 0.828 | 0.732 | 
| 62.50 | 0.496 | 0.400 | 
| 31.25 | 0.372 | 0.276 | 
| 15.63 | 0.238 | 0.142 | 
| 0.00 | 0.096 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 83-96% | 89% | 
| EDTA plasma(n=5) | 92-107% | 99% | 
| Heparin plasma(n=5) | 80-97% | 88% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 86-102% | 87-103% | 85-101% | 81-99% | 
| EDTA plasma(n=5) | 95-104% | 93-105% | 82-101% | 91-103% | 
| Heparin plasma(n=5) | 83-95% | 88-102% | 85-97% | 80-92% |