| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 50.00 | 2.204 | 2.091 | 
| 25.00 | 1.674 | 1.561 | 
| 12.50 | 1.186 | 1.073 | 
| 6.25 | 0.682 | 0.569 | 
| 3.13 | 0.543 | 0.430 | 
| 1.57 | 0.230 | 0.117 | 
| 0.79 | 0.251 | 0.138 | 
| 0.00 | 0.113 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 93-107% | 100% | 
| EDTA plasma(n=5) | 87-99% | 93% | 
| Heparin plasma(n=5) | 92-107% | 99% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 82-93% | 96-105% | 87-98% | 85-97% | 
| EDTA plasma(n=5) | 87-96% | 85-98% | 89-102% | 85-98% | 
| Heparin plasma(n=5) | 86-94% | 85-95% | 89-99% | 85-95% |