| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 2500.00 | 2.058 | 1.971 | 
| 1250.00 | 1.625 | 1.538 | 
| 625.00 | 1.239 | 1.152 | 
| 312.50 | 0.797 | 0.710 | 
| 156.25 | 0.457 | 0.370 | 
| 78.13 | 0.329 | 0.242 | 
| 39.07 | 0.247 | 0.160 | 
| 0.00 | 0.087 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 94-105% | 98% | 
| EDTA plasma(n=5) | 83-97% | 91% | 
| Heparin plasma(n=5) | 82-94% | 88% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 88-97% | 85-101% | 86-98% | 96-107% | 
| EDTA plasma(n=5) | 87-94% | 85-96% | 89-100% | 85-96% | 
| Heparin plasma(n=5) | 87-96% | 85-98% | 89-102% | 85-98% |