| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 5000.00 | 2.103 | 2.021 | 
| 2500.00 | 1.628 | 1.546 | 
| 1250.00 | 1.240 | 1.158 | 
| 625.00 | 0.839 | 0.757 | 
| 312.50 | 0.517 | 0.435 | 
| 156.25 | 0.475 | 0.393 | 
| 78.13 | 0.187 | 0.105 | 
| 0.00 | 0.082 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 80-96% | 88% | 
| EDTA plasma(n=5) | 80-92% | 86% | 
| Heparin plasma(n=5) | 78-95% | 86% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 86-93% | 93-101% | 86-97% | 90-101% | 
| EDTA plasma(n=5) | 85-94% | 87-101% | 89-99% | 85-95% | 
| Heparin plasma(n=5) | 85-94% | 83-96% | 87-98% | 87-100% |