| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 10.00 | 1.941 | 1.868 | 
| 5.00 | 1.643 | 1.570 | 
| 2.50 | 1.144 | 1.071 | 
| 1.25 | 0.969 | 0.896 | 
| 0.63 | 0.543 | 0.470 | 
| 0.32 | 0.324 | 0.251 | 
| 0.16 | 0.142 | 0.069 | 
| 0.00 | 0.073 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 92-106% | 99% | 
| EDTA plasma(n=5) | 88-103% | 95% | 
| Heparin plasma(n=5) | 77-95% | 86% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 86-95% | 85-96% | 89-100% | 85-96% | 
| EDTA plasma(n=5) | 93-106% | 87-99% | 85-93% | 79-96% | 
| Heparin plasma(n=5) | 97-105% | 84-96% | 87-98% | 95-102% |