| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 50.00 | 2.132 | 2.047 | 
| 25.00 | 1.487 | 1.402 | 
| 12.50 | 1.174 | 1.089 | 
| 6.25 | 0.743 | 0.658 | 
| 3.13 | 0.596 | 0.511 | 
| 1.57 | 0.389 | 0.304 | 
| 0.79 | 0.248 | 0.163 | 
| 0.00 | 0.085 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 82-94% | 88% | 
| EDTA plasma(n=5) | 89-107% | 98% | 
| Heparin plasma(n=5) | 85-97% | 91% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 85-94% | 87-96% | 92-101% | 82-98% | 
| EDTA plasma(n=5) | 85-94% | 87-96% | 82-90% | 92-101% | 
| Heparin plasma(n=5) | 85-102% | 83-96% | 95-102% | 87-98% |