| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 100.00 | 2.128 | 2.039 | 
| 50.00 | 1.615 | 1.526 | 
| 25.00 | 1.073 | 0.984 | 
| 12.50 | 0.825 | 0.736 | 
| 6.25 | 0.567 | 0.478 | 
| 3.13 | 0.357 | 0.268 | 
| 1.57 | 0.175 | 0.086 | 
| 0.00 | 0.089 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 95-103% | 99% | 
| EDTA plasma(n=5) | 92-102% | 97% | 
| Heparin plasma(n=5) | 81-95% | 88% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 81-94% | 78-90% | 83-95% | 86-97% | 
| EDTA plasma(n=5) | 87-98% | 92-105% | 83-92% | 89-98% | 
| Heparin plasma(n=5) | 86-98% | 83-96% | 95-104% | 81-95% |