| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 2000.00 | 2.231 | 2.135 | 
| 1000.00 | 1.628 | 1.532 | 
| 500.00 | 1.283 | 1.187 | 
| 250.00 | 0.682 | 0.586 | 
| 125.00 | 0.518 | 0.422 | 
| 62.50 | 0.335 | 0.239 | 
| 31.25 | 0.196 | 0.100 | 
| 0.00 | 0.096 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 80-90% | 85% | 
| EDTA plasma(n=5) | 84-106% | 100% | 
| Heparin plasma(n=5) | 87-93% | 90% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 85-94% | 97-106% | 89-97% | 95-104% | 
| EDTA plasma(n=5) | 92-103% | 83-98% | 88-97% | 96-104% | 
| Heparin plasma(n=5) | 85-97% | 84-100% | 95-103% | 88-104% |