48 T | $412.00 |
96 T | $588.00 |
Concentration (pg/mL) | OD | Corrected OD |
---|---|---|
1000.00 | 2.085 | 1.997 |
500.00 | 1.645 | 1.557 |
250.00 | 1.043 | 0.955 |
125.00 | 0.756 | 0.668 |
62.50 | 0.556 | 0.468 |
31.25 | 0.312 | 0.224 |
15.63 | 0.168 | 0.080 |
0.00 | 0.088 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 87-99% | 93% |
EDTA plasma(n=5) | 85-99% | 92% |
Heparin plasma(n=5) | 78-92% | 85% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 89-102% | 97-103% | 88-96% | 95-101% |
EDTA plasma(n=5) | 85-97% | 90-99% | 82-101% | 93-105% |
Heparin plasma(n=5) | 82-93% | 96-105% | 87-98% | 85-97% |