| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 10.00 | 2.154 | 2.065 | 
| 5.00 | 1.628 | 1.539 | 
| 2.50 | 1.236 | 1.147 | 
| 1.25 | 0.815 | 0.726 | 
| 0.63 | 0.556 | 0.467 | 
| 0.32 | 0.385 | 0.296 | 
| 0.16 | 0.165 | 0.076 | 
| 0.00 | 0.089 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 85-97% | 91% | 
| EDTA plasma(n=5) | 96-109% | 102% | 
| Heparin plasma(n=5) | 81-95% | 88% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 88-103% | 83-94% | 92-105% | 89-97% | 
| EDTA plasma(n=5) | 87-98% | 85-96% | 92-102% | 81-93% | 
| Heparin plasma(n=5) | 88-102% | 87-102% | 85-96% | 87-101% |