| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 1000.00 | 0.197 | |
| 500.00 | 0.332 | |
| 250.00 | 0.567 | |
| 125.00 | 0.878 | |
| 62.50 | 1.157 | |
| 31.25 | 1.592 | |
| 15.63 | 1.697 | |
| 0.00 | 2.213 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 83-95% | 89% | 
| EDTA plasma(n=5) | 87-97% | 90% | 
| Heparin plasma(n=5) | 81-95% | 91% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 95-106% | 83-96% | 87-98% | 95-102% | 
| EDTA plasma(n=5) | 79-97% | 90-99% | 95-103% | 89-97% | 
| Heparin plasma(n=5) | 93-102% | 83-95% | 83-95% | 96-101% |