| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 100.00 | 1.983 | 1.896 | 
| 50.00 | 1.647 | 1.560 | 
| 25.00 | 1.187 | 1.100 | 
| 12.50 | 0.877 | 0.790 | 
| 6.25 | 0.572 | 0.485 | 
| 3.13 | 0.312 | 0.225 | 
| 1.57 | 0.319 | 0.232 | 
| 0.00 | 0.087 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 83-99% | 95% | 
| EDTA plasma(n=5) | 80-96% | 88% | 
| Heparin plasma(n=5) | 87-99% | 93% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 89-99% | 83-96% | 98-104% | 88-101% | 
| EDTA plasma(n=5) | 79-91% | 96-105% | 80-92% | 91-103% | 
| Heparin plasma(n=5) | 92-104% | 87-96% | 89-101% | 82-90% |