| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 1.952 | 1.870 |
| 1000.00 | 1.637 | 1.555 |
| 500.00 | 1.243 | 1.161 |
| 250.00 | 0.937 | 0.855 |
| 125.00 | 0.563 | 0.481 |
| 62.50 | 0.315 | 0.233 |
| 31.25 | 0.135 | 0.053 |
| 0.00 | 0.082 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-99% | 93% |
| EDTA plasma(n=5) | 78-92% | 85% |
| Heparin plasma(n=5) | 77-95% | 86% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 92-102% | 81-93% | 86-98% | 91-103% |
| EDTA plasma(n=5) | 88-97% | 86-92% | 79-103% | 89-97% |
| Heparin plasma(n=5) | 95-104% | 87-96% | 82-96% | 89-103% |