| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 1000.00 | 2.121 | 2.033 | 
| 500.00 | 1.562 | 1.474 | 
| 250.00 | 1.183 | 1.095 | 
| 125.00 | 0.752 | 0.664 | 
| 62.50 | 0.457 | 0.369 | 
| 31.25 | 0.362 | 0.274 | 
| 15.63 | 0.205 | 0.117 | 
| 0.00 | 0.088 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 95-108% | 101% | 
| EDTA plasma(n=5) | 82-95% | 88% | 
| Heparin plasma(n=5) | 80-93% | 86% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 92-102% | 81-93% | 86-98% | 91-103% | 
| EDTA plasma(n=5) | 97-106% | 86-92% | 91-103% | 89-97% | 
| Heparin plasma(n=5) | 85-94% | 97-106% | 92-101% | 82-90% |