| 48 T | $412.00 | 
| 96 T | $588.00 | 
| 96 T*5 | $2499.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 1000.00 | 1.916 | 1.829 | 
| 500.00 | 1.631 | 1.544 | 
| 250.00 | 1.223 | 1.136 | 
| 125.00 | 0.863 | 0.776 | 
| 62.50 | 0.548 | 0.461 | 
| 31.25 | 0.325 | 0.238 | 
| 15.63 | 0.235 | 0.148 | 
| 0.00 | 0.087 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 95-107% | 101% | 
| EDTA plasma(n=5) | 82-95% | 88% | 
| Heparin plasma(n=5) | 85-97% | 91% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 87-96% | 92-105% | 89-102% | 92-101% | 
| EDTA plasma(n=5) | 88-97% | 91-105% | 87-98% | 95-105% | 
| Heparin plasma(n=5) | 84-99% | 83-101% | 86-99% | 93-101% |