| 48 T | $412.00 | 
| 96 T | $588.00 | 
| 96 T*5 | $2499.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 10.00 | 2.013 | 1.918 | 
| 5.00 | 1.562 | 1.467 | 
| 2.50 | 1.139 | 1.044 | 
| 1.25 | 0.797 | 0.702 | 
| 0.63 | 0.507 | 0.412 | 
| 0.32 | 0.302 | 0.207 | 
| 0.16 | 0.215 | 0.120 | 
| 0.00 | 0.095 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 81-103% | 87% | 
| EDTA plasma(n=5) | 93-105% | 99% | 
| Heparin plasma(n=5) | 84-93% | 89% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 85-94% | 87-101% | 86-97% | 88-95% | 
| EDTA plasma(n=5) | 87-93% | 91-103% | 82-99% | 99-104% | 
| Heparin plasma(n=5) | 84-99% | 81-95% | 98-107% | 96-101% |