| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 1.990 | 1.913 |
| 5.00 | 1.643 | 1.566 |
| 2.50 | 1.135 | 1.058 |
| 1.25 | 0.752 | 0.675 |
| 0.63 | 0.527 | 0.450 |
| 0.32 | 0.265 | 0.188 |
| 0.16 | 0.213 | 0.136 |
| 0.00 | 0.077 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 92-102% | 97% |
| EDTA plasma(n=5) | 81-95% | 88% |
| Heparin plasma(n=5) | 95-103% | 99% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 82-93% | 95-104% | 82-96% | 85-98% |
| EDTA plasma(n=5) | 93-102% | 80-93% | 91-100% | 88-95% |
| Heparin plasma(n=5) | 81-94% | 96-105% | 92-103% | 82-96% |