| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 10.00 | 2.154 | 2.086 | 
| 5.00 | 1.565 | 1.497 | 
| 2.50 | 1.205 | 1.137 | 
| 1.25 | 0.872 | 0.804 | 
| 0.63 | 0.427 | 0.359 | 
| 0.32 | 0.302 | 0.234 | 
| 0.16 | 0.226 | 0.158 | 
| 0.00 | 0.068 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 90-105% | 97% | 
| EDTA plasma(n=5) | 95-107% | 101% | 
| Heparin plasma(n=5) | 78-90% | 84% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 88-100% | 95-103% | 89-97% | 85-94% | 
| EDTA plasma(n=5) | 87-98% | 92-104% | 82-96% | 95-102% | 
| Heparin plasma(n=5) | 87-98% | 92-105% | 83-92% | 89-98% |