| 48 T | $320.00 |
| 96 T | $458.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 500.00 | 2.201 | 2.093 |
| 250.00 | 1.611 | 1.503 |
| 125.00 | 1.231 | 1.123 |
| 62.50 | 0.815 | 0.707 |
| 31.25 | 0.567 | 0.459 |
| 15.63 | 0.351 | 0.243 |
| 7.82 | 0.221 | 0.113 |
| 0.00 | 0.108 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 90-105% | 97% |
| EDTA plasma(n=5) | 80-97% | 88% |
| Heparin plasma(n=5) | 78-92% | 85% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-97% | 86-92% | 81-93% | 93-104% |
| EDTA plasma(n=5) | 87-102% | 85-94% | 87-101% | 89-99% |
| Heparin plasma(n=5) | 88-104% | 84-98% | 95-105% | 88-102% |