| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 2.075 | 1.989 |
| 2500.00 | 1.502 | 1.416 |
| 1250.00 | 1.183 | 1.097 |
| 625.00 | 0.877 | 0.791 |
| 312.50 | 0.457 | 0.371 |
| 156.25 | 0.329 | 0.243 |
| 78.13 | 0.207 | 0.121 |
| 0.00 | 0.086 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 95-107% | 101% |
| EDTA plasma(n=5) | 80-93% | 86% |
| Heparin plasma(n=5) | 94-105% | 98% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-99% | 86-92% | 89-103% | 97-105% |
| EDTA plasma(n=5) | 82-96% | 85-96% | 87-101% | 89-97% |
| Heparin plasma(n=5) | 85-101% | 84-98% | 85-93% | 86-98% |