| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 1.997 | 1.916 |
| 2500.00 | 1.586 | 1.505 |
| 1250.00 | 1.223 | 1.142 |
| 625.00 | 0.697 | 0.616 |
| 312.50 | 0.552 | 0.471 |
| 156.25 | 0.322 | 0.241 |
| 78.13 | 0.156 | 0.075 |
| 0.00 | 0.081 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 95-108% | 101% |
| EDTA plasma(n=5) | 82-97% | 89% |
| Heparin plasma(n=5) | 84-93% | 87% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-105% | 86-102% | 97-106% | 89-97% |
| EDTA plasma(n=5) | 95-103% | 88-96% | 81-95% | 91-102% |
| Heparin plasma(n=5) | 89-97% | 86-102% | 94-106% | 85-97% |