| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 1.952 | 1.848 |
| 5.00 | 1.637 | 1.533 |
| 2.50 | 1.125 | 1.021 |
| 1.25 | 0.733 | 0.629 |
| 0.63 | 0.508 | 0.404 |
| 0.32 | 0.369 | 0.265 |
| 0.16 | 0.216 | 0.112 |
| 0.00 | 0.104 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 78-92% | 85% |
| EDTA plasma(n=5) | 92-107% | 99% |
| Heparin plasma(n=5) | 87-99% | 93% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 93-102% | 86-92% | 96-105% | 89-97% |
| EDTA plasma(n=5) | 90-99% | 83-97% | 81-103% | 89-101% |
| Heparin plasma(n=5) | 79-91% | 95-104% | 93-102% | 79-94% |