| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 10.00 | 2.146 | 2.057 | 
| 5.00 | 1.638 | 1.549 | 
| 2.50 | 1.235 | 1.146 | 
| 1.25 | 0.901 | 0.812 | 
| 0.63 | 0.516 | 0.427 | 
| 0.32 | 0.389 | 0.300 | 
| 0.16 | 0.165 | 0.076 | 
| 0.00 | 0.089 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 96-108% | 102% | 
| EDTA plasma(n=5) | 80-95% | 87% | 
| Heparin plasma(n=5) | 85-97% | 91% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 85-96% | 86-95% | 90-102% | 81-90% | 
| EDTA plasma(n=5) | 92-104% | 86-98% | 80-93% | 89-97% | 
| Heparin plasma(n=5) | 85-90% | 83-96% | 95-102% | 88-95% |