| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.058 | 1.972 |
| 500.00 | 1.598 | 1.512 |
| 250.00 | 1.067 | 0.981 |
| 125.00 | 0.837 | 0.751 |
| 62.50 | 0.556 | 0.470 |
| 31.25 | 0.313 | 0.227 |
| 15.63 | 0.100 | 0.014 |
| 0.00 | 0.086 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 81-103% | 87% |
| EDTA plasma(n=5) | 93-105% | 99% |
| Heparin plasma(n=5) | 84-93% | 89% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-100% | 84-99% | 85-97% | 88-102% |
| EDTA plasma(n=5) | 88-103% | 86-97% | 87-102% | 89-101% |
| Heparin plasma(n=5) | 87-98% | 92-101% | 82-98% | 89-97% |