| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 20.00 | 1.952 | 1.863 |
| 10.00 | 1.700 | 1.611 |
| 5.00 | 1.235 | 1.146 |
| 2.50 | 0.877 | 0.788 |
| 1.25 | 0.589 | 0.500 |
| 0.63 | 0.239 | 0.150 |
| 0.32 | 0.175 | 0.086 |
| 0.00 | 0.089 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 84-106% | 100% |
| EDTA plasma(n=5) | 87-93% | 90% |
| Heparin plasma(n=5) | 80-90% | 85% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-92% | 79-96% | 89-100% | 85-96% |
| EDTA plasma(n=5) | 81-94% | 87-105% | 79-95% | 96-105% |
| Heparin plasma(n=5) | 82-94% | 95-102% | 87-103% | 88-95% |