| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.179 | 2.114 |
| 5.00 | 1.506 | 1.441 |
| 2.50 | 1.043 | 0.978 |
| 1.25 | 0.756 | 0.691 |
| 0.63 | 0.518 | 0.453 |
| 0.32 | 0.356 | 0.291 |
| 0.16 | 0.196 | 0.131 |
| 0.00 | 0.065 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 82-94% | 88% |
| EDTA plasma(n=5) | 94-105% | 98% |
| Heparin plasma(n=5) | 83-97% | 91% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 86-97% | 88-95% | 92-101% | 87-98% |
| EDTA plasma(n=5) | 87-102% | 82-91% | 86-93% | 81-94% |
| Heparin plasma(n=5) | 79-96% | 85-99% | 79-93% | 89-97% |