| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 4000.00 | 1.946 | 1.857 |
| 2000.00 | 1.624 | 1.535 |
| 1000.00 | 1.232 | 1.143 |
| 500.00 | 0.837 | 0.748 |
| 250.00 | 0.556 | 0.467 |
| 125.00 | 0.335 | 0.246 |
| 62.50 | 0.213 | 0.124 |
| 0.00 | 0.089 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-97% | 88% |
| EDTA plasma(n=5) | 83-95% | 89% |
| Heparin plasma(n=5) | 81-95% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 79-96% | 85-99% | 79-93% | 89-97% |
| EDTA plasma(n=5) | 86-95% | 93-103% | 85-94% | 87-101% |
| Heparin plasma(n=5) | 91-99% | 83-97% | 87-105% | 87-99% |