| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 150.00 | 2.056 | 1.974 |
| 75.00 | 1.563 | 1.481 |
| 37.50 | 1.067 | 0.985 |
| 18.75 | 0.852 | 0.770 |
| 9.38 | 0.563 | 0.481 |
| 4.69 | 0.315 | 0.233 |
| 2.35 | 0.135 | 0.053 |
| 0.00 | 0.082 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-95% | 90% |
| EDTA plasma(n=5) | 80-95% | 87% |
| Heparin plasma(n=5) | 84-102% | 93% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 78-92% | 96-105% | 86-97% | 85-101% |
| EDTA plasma(n=5) | 92-101% | 87-98% | 85-92% | 79-96% |
| Heparin plasma(n=5) | 87-94% | 85-96% | 89-100% | 85-96% |