| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.037 | 1.944 |
| 5.00 | 1.583 | 1.490 |
| 2.50 | 1.032 | 0.939 |
| 1.25 | 0.837 | 0.744 |
| 0.63 | 0.556 | 0.463 |
| 0.32 | 0.312 | 0.219 |
| 0.16 | 0.196 | 0.103 |
| 0.00 | 0.093 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-95% | 87% |
| EDTA plasma(n=5) | 84-98% | 91% |
| Heparin plasma(n=5) | 79-97% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-97% | 91-105% | 87-98% | 95-105% |
| EDTA plasma(n=5) | 78-94% | 87-96% | 85-93% | 88-103% |
| Heparin plasma(n=5) | 85-92% | 90-96% | 87-1028% | 85-97% |